Bacterial / Viral filter Iso-Gard

Brand: Gibeck

Medium – hydrophobic depth filter made of polypropylene, electrostatically charged
Tidal volume – 150–1000 ml
Connectors – 15M/22F–22M/15F
How to use:

  • Place the filter in a vertical position
  • Tube (breathing circuit) should be at the level lower than the filter
  • Connect the gas supply tube
  • Connect and disconnect the filter by turning it

Cautions:

  • Condensate may accumulate on the patient side of the product. To prevent accumulation of condensate, position filter to allow water to drain away from upstream side of the filter.
  • Take total system dead space into consideration if used as an HME.

Warnings:

  • Do not soak, rinse, wash or sterilize (gas, steam or cold) this product.
  • Discard after use.
  • Reuse of medical devices intended for single use can cause a loss of functional capacity and reduced efficiency. Reusing disposable products also leads to viral, bacterial, fungal and other opportunistic complications. Existing methods of cleaning and sterilization, as well as re-treatment are not applicable on these products. The product is not a subject to cleaning, disinfection or sterilization
  • Replace filter immediately if there is increasing resistance or any suspicion of contamination, occlusion or other indications of malfunction present.

Contraindications:
Make sure if you use drugs with surfactant properties that the hydrophobic pores are not blocked This can happen if drugs are nebulized in a circuit with a cascade humidifier. Always consult with your drug manufacturer before nebulization of drugs or solutions for the first time.
Storage conditions
In a dry, dark place, at the temperature of 0C + 25C